Products List
On-line+
- HNPBL-1000
- HNPDM-1802
- HNPJF-1500B
- HNPRT-18B
- HNPRX-1000B
- HNPRX-1000C
- HNPSE1000
Portable+
- A-B SF6
Contact Us
Products Name:
HNPSP-10 Strap Type Purity Analyzer

Overview
The HNPSP-10 strap type purity analyzer has reasonable structure, patented gas chamber and dual temperature compensation thermal conductivity sensor, and its measurement results are not affected by the ambient temperature.
This device has higher precision in the field experiments because it is not affected by the ambient temperature. The strap type design also meets the field operation needs.
Main Features
Dual temperature compensation thermal conductivity sensor
Patented gas chamber for temperature compensation
Strap type design
Measurement results not affected by the ambient temperature
Built-in pressure stabilizing valve and electronic mass flowmeter
No warm-up waiting, and low gas consumption
Technical Parameters
Measurement range | 0~100%, typical precision range 90~100% |
Precision | ±0.6% |
Response time | 0.1% |
Flow rate of sample gas | 300~500mL/min |
Pressure of sample gas | ≤1.0Mpa |
Working power supply | AC 220V/50Hz or built-in lithium battery |
Working environment | Ambient temperature: -30°C~60°C |
Relative humidity: 0~90%RH | |
Storage capacity | ≮200 normal data |
Weight | About 4kg |
Dimensions | About 260mm×195mm×125mm |
Usage time | ≮8 hours |